20 questions, each with a clear explanation.

Quick PracticeQuestion 1 of 20
Your progress through this set
Safety, Compliance, and Coordinated Patient CareStep 1 · Choose one answer

A coworker from the billing office stops by the EKG room and asks to see the 12-lead tracing you just recorded for a well-known local coach. The coworker has no role in the coach's care and says it is just curiosity. What should you do?

Question review

All 20 NHA CET questions with explanations

Try the set above first, then use this list to review. Each answer and explanation stays closed until you open it.

  1. Question 1 · Safety, Compliance, and Coordinated Patient Care

    A coworker from the billing office stops by the EKG room and asks to see the 12-lead tracing you just recorded for a well-known local coach. The coworker has no role in the coach's care and says it is just curiosity. What should you do?

    1. Show the tracing, since everyone on staff is already bound by HIPAA
    2. Decline, since access is limited to staff who need it for their job
    3. Show only the rhythm strip, with the patient's name and ID covered
    4. Share it later, once the provider has reviewed and signed the report
    Show answer and explanation for question 1

    Correct answer: B · Decline, since access is limited to staff who need it for their job

    Federal HIPAA rules (45 CFR 164.514(d)) require a covered entity to identify which workforce members need access to protected health information to do their jobs and to limit access to that need. A billing coworker with no role in this patient's care has no work reason to see the tracing, and curiosity is not one, so you decline. Being bound by HIPAA does not create a right to look. Covering the name and ID does not help, because the coworker already knows whose tracing it is. The provider's review changes nothing about who may see the record.

    Look for this next time: Access to a patient's EKG is limited to staff who need it for their work, and curiosity is never a reason.

  2. Question 2 · Safety, Compliance, and Coordinated Patient Care

    You have just finished a resting 12-lead EKG on a patient with intact skin. The machine uses reusable lead wires and limb clips that touched only the patient's skin and the electrodes. Your facility's policy calls for processing them between patients. Which level of processing fits?

    1. Steam sterilization in the facility's central processing department
    2. High-level disinfection, as used for scopes that enter the body
    3. A rinse under running tap water, then air-drying on a towel
    4. Cleaning, then a low-level disinfectant used as its label directs
    Show answer and explanation for question 2

    Correct answer: D · Cleaning, then a low-level disinfectant used as its label directs

    CDC's disinfection guideline classifies items that touch intact skin but not mucous membranes as noncritical, with examples such as blood pressure cuffs, and lists low-level disinfectants for them. Lead wires and clips used on intact skin fit that group, so they are cleaned and then disinfected with a low-level product, following the label's directions, such as contact time, as federal law requires for EPA-registered disinfectants. Sterilization is for critical items that enter sterile tissue or the bloodstream. High-level disinfection is for semicritical items that touch mucous membranes, such as endoscopes. A tap-water rinse is not disinfection at all.

    Look for this next time: Lead wires and clips that touch only intact skin are noncritical items: clean them, then use a low-level disinfectant as its label directs.

  3. Question 3 · Safety, Compliance, and Coordinated Patient Care

    A patient scheduled for an exercise stress test tomorrow calls the cardiology office: "Should I skip my morning heart pill? A friend said it can change the results." What is the best response from the EKG technician?

    1. Ask the provider who ordered the test before you change any medicine
    2. Skip it, because some medicines can interfere with stress test results
    3. Take half the usual dose as a middle ground between the two choices
    4. Take it as usual, because medicines do not affect stress test results
    Show answer and explanation for question 3

    Correct answer: A · Ask the provider who ordered the test before you change any medicine

    MedlinePlus tells stress test patients to ask their provider whether to take their regular medicines on the test day, because some medicines may interfere with the results, and never to stop a medicine without first talking to the provider. Deciding which medicine to hold is outside an EKG technician's scope of practice, so the technician sends the question to the ordering provider. Telling the patient to skip the pill gets the first half of the fact right but makes a clinical decision. Halving the dose is prescribing. Saying medicines never affect the results is false.

    Look for this next time: Medication questions before a stress test go to the ordering provider; the technician does not decide what to hold.

  4. Question 4 · Safety, Compliance, and Coordinated Patient Care

    A patient with limited English proficiency arrives for a resting EKG with her 12-year-old son, who offers to interpret. The clinic has a qualified phone interpreter service, and nothing about the visit is urgent. How should you explain the procedure to her?

    1. Let her son interpret, since it is a quick test
    2. Skip the explanation and point to show her where she should lie down
    3. Call the clinic's phone interpreter and explain the test through them
    4. Ask her to come back another day with an adult relative who can help her
    Show answer and explanation for question 4

    Correct answer: C · Call the clinic's phone interpreter and explain the test through them

    Under the federal Section 1557 rule (45 CFR 92.201), a covered health program must offer a qualified interpreter when interpretation is needed and must not rely on a minor child to interpret, except briefly in an emergency while a qualified interpreter is found. Here the clinic has an interpreter service and nothing is urgent, so you use it to explain the test and answer her questions. Letting her son interpret is the situation the rule restricts. Pointing instead of explaining leaves her unable to understand the test. Sending her away to bring her own interpreter is also barred, because the rule says a covered program must not require a patient to provide one.

    Look for this next time: Use a qualified interpreter for a patient with limited English proficiency; a child interprets only as a stopgap in an emergency.

  5. Question 5 · Safety, Compliance, and Coordinated Patient Care

    You have just attached a 24-hour Holter monitor to a patient who has felt his heart flutter several times this week. Which instruction matters most for matching his symptoms to the recording?

    1. Keep a diary of activities and how you feel, noting when each occurs
    2. Stay in bed for 24 hours to keep the recording clean
    3. Take the monitor off at bedtime and put it back on in the morning
    4. Shower as usual, because the monitor and its wires are all waterproof
    Show answer and explanation for question 5

    Correct answer: A · Keep a diary of activities and how you feel, noting when each occurs

    MedlinePlus tells Holter patients to keep a diary of their activities and how they feel, and to record symptoms accurately so the provider can match them with the monitor findings; noting when each symptom happens is what makes that match possible. Staying in bed defeats the purpose, since patients are told to continue their normal activities. Taking the monitor off at night leaves gaps, and patients are told to keep it close to the body. A Holter monitor must not get wet, which is why patients shower or bathe before it is attached.

    Look for this next time: A Holter diary of activities and symptoms, with times, lets the provider link what the patient felt to the tracing.

  6. Question 6 · Safety, Compliance, and Coordinated Patient Care

    You are called to record a STAT 12-lead EKG in a busy emergency department. The machine's screen still shows the previous patient's name and date of birth. What should you do before recording and transmitting the tracing?

    1. Record first and fix the patient name in the chart afterward
    2. Confirm two identifiers with the patient and enter them in the machine
    3. Match the room number on the order, since the patient is already in bed
    4. Keep the stored details, since the tracing will be read right away anyway
    Show answer and explanation for question 6

    Correct answer: B · Confirm two identifiers with the patient and enter them in the machine

    The Society for Cardiological Science and Technology's recording guideline calls for identifying every patient with at least two unique identifiers, such as name and date of birth. It warns that some machines keep the last patient's details, which then print on the new tracing if no one changes them, so the details must be confirmed and entered before recording. Recording first and fixing the record later lets a STAT tracing reach the provider under the wrong name, which is the error this check prevents. A room number identifies a bed, not a person. Keeping the stored details guarantees the mismatch.

    Look for this next time: Confirm two patient identifiers and enter them before every tracing, because a machine can carry over the last patient's details.

  7. Question 7 · EKG Acquisition

    You are placing chest electrodes for a resting 12-lead EKG. You find the angle of Louis, slide your finger to the patient's right, and locate the second intercostal space. Where does the V1 electrode go?

    1. Second intercostal space at the right sternal edge
    2. Fifth intercostal space, right midclavicular line
    3. Fourth intercostal space at the left sternal edge
    4. Fourth intercostal space at the right sternal edge
    Show answer and explanation for question 7

    Correct answer: D · Fourth intercostal space at the right sternal edge

    The SCST recording guideline places V1 in the fourth intercostal space at the right sternal edge and V2 in the fourth space at the left sternal edge. To find it, count down from the second intercostal space beside the angle of Louis to the fourth, then slide toward the sternum. The guideline notes that V1 and V2 are often placed too high, which is the second-space error here. The fifth intercostal space at the right midclavicular line mirrors V4 on the right side and is used for the right-sided lead V4R, not for V1. The fourth space at the left sternal edge is V2.

    Look for this next time: V1 goes in the fourth intercostal space at the right sternal edge; count the spaces down from the angle of Louis.

  8. Question 8 · EKG Acquisition

    After placing V4 in the fifth intercostal space at the midclavicular line, a student places V5 and V6 by following the curve of the fifth intercostal space toward the armpit instead of staying level with V4. Where should V5 and V6 go?

    1. V5 at the midaxillary line and V6 at the posterior axillary line, in the fifth space
    2. V5 and V6 both in the fifth intercostal space, wherever it runs toward the armpit
    3. V5 at the anterior axillary line and V6 at the midaxillary line, level with V4
    4. V5 and V6 both in the sixth intercostal space, one full space below V4
    Show answer and explanation for question 8

    Correct answer: C · V5 at the anterior axillary line and V6 at the midaxillary line, level with V4

    In the SCST recording guideline, V5 goes on the left anterior axillary line and V6 on the left midaxillary line, both at the same horizontal level as V4, which sits in the fifth intercostal space at the midclavicular line. Following the intercostal space toward the armpit moves the electrodes off that level, and the guideline notes that misplaced chest electrodes can produce misleading waveform changes. Moving each electrode one line back puts V6 on the posterior axillary line, the position of the posterior lead V7. Dropping both to the sixth space places them lower than the standard level.

    Look for this next time: V5 and V6 stay at the same horizontal level as V4, on the anterior axillary and midaxillary lines.

  9. Question 9 · EKG Acquisition

    For a resting diagnostic 12-lead EKG, a coworker suggests placing the four limb electrodes on the shoulders and lower abdomen, the way telemetry patches are placed, to save time. What is the standard placement for this test?

    1. Shoulders and lower abdomen, because limb positions are interchangeable
    2. Forearms and lower legs, just above the wrists and the ankles
    3. Upper arms and upper thighs, as close to the torso as possible
    4. Chest wall near each shoulder and over each hip bone
    Show answer and explanation for question 9

    Correct answer: B · Forearms and lower legs, just above the wrists and the ankles

    The SCST recording guideline places limb electrodes on the forearms and lower legs, just above the wrists and ankles, whenever possible. Moving them up the limbs can change the tracing and is avoided unless there is a significant tremor or an amputation, and limb electrodes must not go on the torso for a 12-lead, because that significantly changes wave amplitudes and can invalidate the recording for diagnosis. Telemetry serves continuous monitoring, a different purpose from a diagnostic 12-lead, so its patch positions do not carry over. Upper arms and thighs, or the chest wall near the shoulders and hips, are the nonstandard positions the guideline warns against.

    Look for this next time: For a diagnostic 12-lead, limb electrodes go on the forearms and lower legs, never on the torso.

  10. Question 10 · EKG Acquisition

    A routine 12-lead EKG on an adult whose earlier tracings were normal shows negative P waves and negative QRS complexes in lead I. Before the tracing goes to the provider, what should the technician check first?

    1. Whether the right and left arm lead wires were switched
    2. Whether the paper speed was set to 50 mm per second
    3. Whether the chest electrodes were placed one space too low
    4. Whether the patient moved or talked during the recording
    Show answer and explanation for question 10

    Correct answer: A · Whether the right and left arm lead wires were switched

    Lead I records between the right arm (negative) and left arm (positive) electrodes, so switching the two arm wires turns lead I upside down. The SCST recording guideline says dextrocardia may be suspected when lead I shows negative P waves and QRS complexes, but only in the absence of a technical error such as reversed right and left arm connections, so the reversal is ruled out first and the tracing repeated with correct connections. A faster paper speed widens the waveforms without inverting them. Chest electrodes do not form lead I. Movement adds artifact, not a clean inversion.

    Look for this next time: Negative P waves and QRS complexes in lead I call for a check of the arm lead wires before anything else.

  11. Question 11 · EKG Acquisition

    On a 12-lead EKG recorded at 25 mm/s and 10 mm/mV, the QRS complexes in several chest leads are so tall that they run into the leads above and below. What is an accepted way to get a readable recording?

    1. Switch the paper speed to 12.5 mm/s and leave the gain at 10 mm/mV
    2. Turn on the machine's muscle filter and print the tracing again
    3. Move the chest electrodes one intercostal space lower and re-record
    4. Re-record at a gain of 5 mm/mV, clearly marking the change on it
    Show answer and explanation for question 11

    Correct answer: D · Re-record at a gain of 5 mm/mV, clearly marking the change on it

    The SCST recording guideline sets the standard at 25 mm/s and 10 mm/mV and says that when complexes are tall enough to overlap, the gain may be lowered to 5 mm/mV for clearer viewing and measurement, with any gain change clearly marked on the ECG. Halving the gain halves the height of every waveform, so the complexes fit. Changing the paper speed stretches or squeezes the tracing in time and does nothing to height. The filter reduces interference but also distorts the ECG, and it has no role in amplitude. Moving the chest electrodes creates a nonstandard tracing.

    Look for this next time: Overlapping tall complexes call for a 5 mm/mV gain, clearly marked; paper speed changes width, not height.

  12. Question 12 · EKG Acquisition

    A nervous patient lies with clenched fists and stiff arms, and the first 12-lead tracing shows irregular, jagged interference across the limb leads. What should the technician do first?

    1. Turn on the machine's muscle filter and repeat the recording
    2. Print the tracing as it is and write 'artifact' across the top
    3. Move the limb electrodes to the shoulders, where the arms are still
    4. Help the patient relax and support the arms, then record again
    Show answer and explanation for question 12

    Correct answer: D · Help the patient relax and support the arms, then record again

    The SCST recording guideline says the patient should be relaxed and comfortable with the limbs supported, because tension records muscle potentials along with cardiac activity, and it notes that clenched fists or stiff arms make a high-quality ECG impossible. The filter comes later: it may be switched on only if muscle interference remains after these efforts, and its use must be marked, because it also distorts the ECG. Printing a tracing full of artifact skips the fix. Moving limb electrodes onto the torso significantly alters wave amplitudes on a 12-lead.

    Look for this next time: For somatic tremor, relax the patient and support the limbs before you consider a filter.

  13. Question 13 · EKG Acquisition

    The chest electrodes keep lifting at the edges on a patient who applied body lotion this morning, and the baseline wanders. What is the best next step?

    1. Press each electrode down harder and tape over all the lifting edges
    2. Spread extra gel over each electrode to improve the contact
    3. Clean and dry the sites, then apply a fresh set of electrodes
    4. Move each electrode to whatever nearby spot it will stick to
    Show answer and explanation for question 13

    Correct answer: C · Clean and dry the sites, then apply a fresh set of electrodes

    The SCST recording guideline names skin preparation as the way to lower skin-to-electrode impedance and get an artifact-free ECG: cleanse the skin with mild soap and water or an alcohol-based wipe, use very light abrasion if needed, and check that electrodes are in good condition with gel that has not dried out. It also says the baseline should be stable, not wandering. The lotion under the electrodes is the problem, so it has to come off before fresh electrodes go on. Pressing harder or taping leaves the lotion in place, and extra gel does not remove it either. Moving electrodes off their landmarks creates a nonstandard tracing.

    Look for this next time: Lotion or oil under the electrodes calls for cleaning and drying the skin, then applying fresh electrodes.

  14. Question 14 · EKG Acquisition

    A patient with heart failure becomes short of breath whenever the bed is flat. The order is for a resting 12-lead EKG. How should you position the patient?

    1. Raise the head of the bed and note that position on the tracing
    2. Keep the bed flat, since a resting EKG must be recorded supine
    3. Have the patient stand, since breathing is easiest when upright
    4. Postpone the EKG until the patient can lie flat without trouble
    Show answer and explanation for question 14

    Correct answer: A · Raise the head of the bed and note that position on the tracing

    The SCST recording guideline notes that many patients are uncomfortable lying flat, that a significant change from the usual position should be documented on the ECG, and that it found no evidence that inclining the patient anywhere between flat and 45 degrees significantly affects the tracing. Raising the head of the bed lets this patient breathe and still gives a usable resting 12-lead, as long as the position is noted. NHA's test plan lists positioning for patients with respiratory issues among the knowledge an EKG technician needs. A flat bed ignores the patient's distress. A tracing recorded upright can differ significantly from one recorded lying down. Postponing delays an ordered test without need.

    Look for this next time: If a patient cannot lie flat, raise the head of the bed and document the position on the tracing.

  15. Question 15 · EKG Acquisition

    During the third stage of a treadmill stress test, the patient says his chest pain is getting worse, and his latest blood pressure is lower than the one before. What should the EKG technician do?

    1. Encourage him to keep walking until he reaches his target heart rate
    2. Alert the supervising provider now, since these are reasons to stop
    3. Raise the incline to reach the target heart rate sooner
    4. Note the symptoms now and report them after the recovery period
    Show answer and explanation for question 15

    Correct answer: B · Alert the supervising provider now, since these are reasons to stop

    MedlinePlus explains that a stress test continues until the patient reaches a target heart rate, develops chest pain or a concerning change in blood pressure, shows EKG changes suggesting the heart muscle is not getting enough oxygen, or has symptoms that keep them from continuing. Worsening chest pain with a falling blood pressure meets that description, so the supervising provider must know right away. NHA's test plan has the EKG technician assist in monitoring the patient and support the response to complications during stress testing. Pushing toward the target rate, or raising the incline, adds strain. Waiting until recovery delays the response.

    Look for this next time: Worsening chest pain or a falling blood pressure during a stress test goes to the supervising provider immediately.

  16. Question 16 · EKG Analysis and Interpretation

    A rhythm strip recorded at 25 mm/s is irregular. You count 9 QRS complexes in a 6-second section of the strip. What is the patient's heart rate?

    1. About 54 per minute
    2. About 45 per minute
    3. About 90 per minute
    4. About 180 per minute
    Show answer and explanation for question 16

    Correct answer: C · About 90 per minute

    The 6-second method counts the QRS complexes in 6 seconds of tracing and multiplies by 10, because 6 seconds times 10 is one minute: 9 × 10 = 90 per minute. At the standard 25 mm/s, 6 seconds is 150 mm of paper, or 30 large 5-mm boxes. NHA's test plan names the 6-second method among the ways to calculate heart rate, and it suits an irregular rhythm, where a single R-R interval gives a different rate for each pair of beats. Multiplying by 6 gives 54, multiplying by 5 gives 45, and 180 doubles the count as if the section were only 3 seconds long.

    Look for this next time: For an irregular rhythm, count the QRS complexes in 6 seconds and multiply by 10.

  17. Question 17 · EKG Analysis and Interpretation

    A rhythm strip shows an irregularly irregular rhythm. Between the QRS complexes, the baseline shows fine, uneven undulations and no distinct P waves. Which rhythm does this describe?

    1. Atrial fibrillation (AF)
    2. Atrial flutter with 2:1 conduction
    3. Sinus bradycardia
    4. First-degree AV block
    Show answer and explanation for question 17

    Correct answer: A · Atrial fibrillation (AF)

    The University of Utah's ECG Learning Center describes atrial fibrillation as poorly defined atrial activity, with coarse or fine undulations or none at all, and an irregularly irregular ventricular response. Both features are in this strip. Atrial flutter shows regular, saw-tooth flutter waves, and with 2:1 conduction the ventricular rhythm is usually regular. Sinus bradycardia is a slow sinus rhythm, and a rhythm from the sinus node has upright P waves in leads I and II. First-degree AV block keeps a P wave before every QRS, only with a PR interval longer than 0.20 second.

    Look for this next time: Irregularly irregular QRS complexes with no distinct P waves point to atrial fibrillation.

  18. Question 18 · EKG Analysis and Interpretation

    On a tracing recorded at 25 mm/s, every P wave is followed by a QRS complex. The PR interval measures 6 small boxes on every beat, and the rhythm is regular. What does this finding suggest?

    1. A normal PR interval for an adult
    2. Second-degree AV block, type I
    3. A short PR interval from preexcitation
    4. First-degree AV block, a prolonged PR
    Show answer and explanation for question 18

    Correct answer: D · First-degree AV block, a prolonged PR

    At 25 mm/s each small box is 1 mm, or 0.04 second, so 6 small boxes are 0.24 second. The University of Utah's ECG Learning Center gives a normal PR interval of 0.12 to 0.20 second and lists first-degree AV block, with a usually constant PR, under a prolonged PR above 0.20 second. In second-degree AV block type I, the PR lengthens until a P wave is not followed by a QRS, but here every P wave conducts with the same PR. A short PR is under 0.12 second, the opposite of this finding. A PR of 0.24 second is above the normal range.

    Look for this next time: A constant PR interval over 0.20 second, with every P wave conducted, is first-degree AV block.

  19. Question 19 · EKG Analysis and Interpretation

    A monitored patient's tracing suddenly shows a chaotic, irregular waveform with no recognizable QRS complexes. When you reach her, she does not respond to tapping and shouting, and she is not breathing normally. What should you do?

    1. Check every lead connection and print a strip before anything else
    2. Shout for help to activate the emergency response, then start CPR
    3. Wait five minutes, then recheck her pulse and breathing
    4. Turn her onto her side and wait with her for the provider to come
    Show answer and explanation for question 19

    Correct answer: B · Shout for help to activate the emergency response, then start CPR

    MedlinePlus's adult CPR page, which cites the American Heart Association's 2025 guideline highlights, says to check for responsiveness and, if there is no response, to shout for help and have someone call for emergency help, begin chest compressions, and use an AED as soon as one is available. A patient who is unresponsive and not breathing normally needs that response now, whatever the monitor shows. NHA's test plan lists action on life-threatening arrhythmias, such as ventricular fibrillation, and names CPR and basic life support as core knowledge. Checking leads is for an alert patient whose tracing looks wrong. Waiting five minutes is longer than the 4 minutes without oxygen after which, MedlinePlus notes, permanent brain damage begins. Turning her and waiting delays compressions.

    Look for this next time: An unresponsive patient who is not breathing normally needs help called and CPR started, whatever the monitor shows.

  20. Question 20 · EKG Analysis and Interpretation

    A patient waiting for a routine pre-op EKG mentions chest pressure. The 12-lead shows ST segments with straight or convex upward elevation in leads II, III, and aVF. What should the technician do?

    1. File the tracing for review later with the other pre-op EKGs
    2. Tell the patient the tracing shows that a heart attack is underway
    3. Bring the tracing and the symptom to the provider immediately
    4. Repeat the EKG after a 30-minute rest
    Show answer and explanation for question 20

    Correct answer: C · Bring the tracing and the symptom to the provider immediately

    The University of Utah's ECG Learning Center describes convex or straight upward ST elevation, for example in leads II, III, and aVF, as abnormal and suggestive of transmural injury or infarction. The SCST recording guideline says features that might need urgent medical attention should be brought to appropriate staff, and that symptoms such as chest pain at the time of recording should be noted on the ECG. That makes this tracing urgent, not routine. Filing it for later review delays care. Telling the patient what the tracing means is a diagnosis, which is the provider's role. Waiting 30 minutes delays an urgent finding.

    Look for this next time: ST elevation with chest symptoms goes to the provider immediately; the technician reports it but does not diagnose.

Use the sample well

The rationale matters more than the score.

A short set can show you what deserves another look. Twenty questions are a small sample, so this set cannot measure readiness for the 100 scored items on the NHA CET exam or predict a score of 390 out of 500.

01

Answer before reviewing

Picture the patient, the electrodes, and the tracing in front of you, choose the one action or reading the scenario calls for, then read why each of the other three options falls short.

02

Read every rationale

Confirm why the best option fits, even when you were right.

03

Follow the pattern

Use recurring misses—not one score—to choose your next focus.